It is mainly used for the creation of anastomosis and the closure of stumps or incisions in digestive tract reconstruction and other organ resection surgeries, and is used in open surgery.
It is mainly used for the creation of anastomosis and the closure of stumps or incisions in digestive tract reconstruction and other organ resection surgeries, and is used in open surgery.
【Scope of application】
Disposable linear cutting staplers and reload is applicable for transaction, resection and the creation of anastomoses for digestive tissue and organ.
【Product performance】
1. The stapler should have good anastomosis and cutting performance. Replace the reloads and make no less than 5 times cut and anastomosis. After each anastomosis, the staples are shaped as "B".
2. The cutting edge after each anastomosis should be neat and free of burrs.
3. The length of each anastomosis line is at least 1.5 times longer than the length of the cutting line.
Basic size of stapler and reloads Unit: mm
|
Size Model |
L |
limit deviation |
L1 |
limit deviation |
H |
limit deviation |
staples |
|
GCLCA-55H |
222 |
±30 |
55 |
±2.0 |
4.3 |
±0.2 |
56 |
|
GCLCA-55M |
222 |
55 |
3.8 |
56 |
|||
|
GCLCA-55L |
222 |
55 |
2.5 |
56 |
|||
|
GCLCA-75H |
272 |
75 |
4.3 |
76 |
|||
|
GCLCA-75M |
272 |
75 |
3.8 |
76 |
|||
|
GCLCA-75L |
272 |
75 |
2.5 |
76 |
|||
|
GCLCA-100H |
322 |
100 |
4.3 |
100 |
|||
|
GCLCA-100M |
322 |
100 |
3.8 |
100 |
|||
|
GCLCA-100L |
322 |
100 |
2.5 |
100 |
Basic reload size Unit: mm
|
Size Model |
L1 |
limit deviation |
H |
limit deviation |
Staple(s) |
|
GCLCA-ZJ55H |
55 |
±2.0 |
4.3 |
±0.2 |
56 |
|
GCLCA-ZJ55M |
55 |
3.8 |
56 |
||
|
GCLCA-ZJ55L |
55 |
2.5 |
56 |
||
|
GCLCA-ZJ75H |
75 |
4.3 |
76 |
||
|
GCLCA-ZJ75M |
75 |
3.8 |
76 |
||
|
GCLCA-ZJ75L |
75 |
2.5 |
76 |
||
|
GCLCA-ZJ100H |
100 |
4.3 |
100 |
||
|
GCLCA-ZJ100M |
100 |
3.8 |
100 |
||
|
GCLCA-ZJ100L |
100 |
2.5 |
100 |
【Product model/specification】
Model specifications of stapler and reloads: GCLCA-55H,GCLCA-55M,GCLCA-55L,GCLCA-75H,GCLCA-75M, GCLCA-75L,GCLCA-100H,GCLCA-100M,GCLCA-100L.
Reload specifications and models:GCLCA-ZJ55H,GCLCA-ZJ55M,GCLCA-ZJ55L, GCLCA-ZJ75H,GCLCA-ZJ75M,GCLCA-ZJ75L,GCLCA-ZJ100H,GCLCA-ZJ100M, GCLCA-ZJ100L.
ZJ represents reloads; 55, 75 and 100 represents reload specifications. H, M, and L represents high nails, medium nails and low nails.
【Instructions for use】 (Not as a technical guideline for surgical anastomosis)
Reload replacement:
(Each stapler can replace up to five reloads in the same operation)
Make sure that when the instrument is fully opened, pull out the fired reloads,and insert a new sterile reload of the corresponding specification into the instrument (a slight click sound during insertion indicates that it is installed in place).
Factory Area (m² )